Q-00028533
Follow up to WO-01276361
Inspection #: 27864
Asset ID: 13245451
Mindray - DPM-6
Serial #: DT-51007737
Notes: Failed Calibration Follow-up required. Recommend remove from service. Reviewed service with customer, quote needed Equipment will not calibrate NOT ABLE TO OBTAIN ANY ACCURATE READINGS WITHIN THE ACCEPTABLE TOLERANCES FOR NIBP AND SPO2
This unit will need to be shipped to our repair depot for Calibration / Repair / Full PM service.
Recommend based on age of unit:
1x- B11775-Mindray DPM6
https://www.ebiotrack.com/workorder.php?wo=66103
Parts Pending1 parts order(s) awaiting fulfillment
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Assigned
0.0d / 0.5d SLA
Awaiting Parts
1.7d / 3.0d SLA
Reschedule
7.0d / 2.0d SLA
On Hold
66.7d
Closed
2.9d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
8/14-RS: e2a Sent with quote and RL
On-Hold Awaiting equipment - Close by EOM and bill for expenses used if no response
6/25-: On-Hold awaiting equipment - box and Packing material delivered - awaiting shipment to depot
🔧Work Performed
Box and packing material - 24 x 18 x 12" Corrugated Boxes
Shipment Tracking Number: 1Z01E3E70397875108
Delivered: 6/19/2025
RMA Tracking Number
1Z01E3E79093785102
Shipper created a label, UPS has not received the package yet.
🔍Technician Findings
depot service requested - quoted - Box and return labels sent to site
⚠️Problem Description
Q-00028533
Follow up to WO-01276361
Inspection #: 27864
Asset ID: 13245451
Mindray - DPM-6
Serial #: DT-51007737
Notes: Failed Calibration Follow-up required. Recommend remove from service. Reviewed service with customer, quote needed Equipment will not calibrate NOT ABLE TO OBTAIN ANY ACCURATE READINGS WITHIN THE ACCEPTABLE TOLERANCES FOR NIBP AND SPO2
This unit will need to be shipped to our repair depot for Calibration / Repair / Full PM service.
Recommend based on age of unit:
1x- B11775-Mindray DPM6
https://www.ebiotrack.com/workorder.php?wo=66103
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Jun 14, 2025
To: bradley.cardenas@unitedveincenters.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01279235Location: Memorial OBLAddress: 11221 Katy Freeway, Houston TX 77079 Houston, TX 77079Device: Problem Description: Q-00028533
Follow up to WO-01276361
Inspection
PDF
ShippingLabel_0001_1Z01E3E79093785102.pdf
Aug 14, 2025
PDF
EMSAR Quote Q-00028533 - United Vein and Vascular Centers 06-13-2025.pdf
Aug 14, 2025
Timeline
📞
Case Linked Jun 4, 3:12 PM
Case Number:00498931
Subject:Needs quote
Status:Closed
⚠️8.0 business days
📋
▶
3
Work Order Created Jun 13, 1:57 PM
Work Order Created → Technician Assigned → Status: Assigned
👷
Technician Assigned by Sherry Reich
Technician:Ryan Schwochow
👷
Status: Assigned by Sherry Reich
From:Entered
To:Assigned
Duration in Previous:1m
📦
▶
2
Parts Requested Jun 13, 2:12 PM
Parts Requested → Status: Awaiting Parts
📦
Status: Awaiting Parts by Ryan Schwochow
From:Assigned
To:Awaiting Parts
Duration in Previous:14m
⏳2.0 business days
🔩
▶
4
2x Parts Added Jun 16, 10:18 AM
2x Parts Added → Status: Reschedule → Parts Requested → Expenses Added
🔄
Status: Reschedule by Emily Wilson
From:Awaiting Parts
To:Reschedule
Duration in Previous:2d 20h
📦
Parts Requested
Order:00076772
Status:Open
💵
Expenses Added
Type:Expenses
Status:Open
Qty:1
⚠️7.0 business days
⏸️
Status: On Hold by Ryan SchwochowJun 25, 2:38 PM
From:Reschedule
To:On Hold
Duration in Previous:9d 4h
🚨67.0 business days
🏁
Status: Closed by Ryan SchwochowSep 29, 12:36 PM
From:On Hold
To:Closed
Duration in Previous:95d 21h
⏳3.0 business days
📄
▶
2
Status: Ready to Bill Oct 1, 4:29 PM
Status: Ready to Bill → Status: Invoiced
💰
Status: Invoiced by Nancy Suarez
From:Invoice Pending
To:Invoiced
Duration in Previous:1m
🧾
Invoice Created Oct 2, 11:00 AM
Invoice Number:SM-242579
Bottlenecks
Reschedule
Duration: 7.0 business days (threshold: 2.0 business days)
Exceeded by 5.0 business days
Closed
Duration: 2.9 business days (threshold: 1.0 business days)