Case 118814- 087 - SN 812868 - UC Health - Calf Actuator - CB - HC - DISPATCH REQUEST ** 1 of 3 at same location
Closed
Model: 087
SN: 812867
Description of Equipment Issue(s) and Service(s) Requested:
Case 122415 – I will need to speak w FT before they go onsite. The customer has a damaged stirrups that will need to
be replaced. Parts tracking # is 884272432392. I do not want the damaged parts back. See attached work instructions.
Sign off sheet will need to be sent to MPI at the time of repair completion. Thank you!
Part Number(s), Descriptions, and Quantities of Parts Suggested and/or Shipped (Include Ship Dates & Tracking Numbers)
Teri Masoner, 816-591-5567 cell, tmasoner@medicalpositioning.com.
***Products can be viewed at wwwmedicalpositioning.com***
Primary Delay5.0 days stuck in "07e-Partial Project"
⏱️
Total Delay2.8 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
0.1d / 1.0d SLA
Assigned
2.0d / 0.5d SLA
Scheduled
1.5d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Completed
0.3d / 1.0d SLA
Closed
2.3d / 1.0d SLA
07e-Partial Project
5.0d
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
9/18/25 NSuarez placed in Partial project awaiting for Parent and child to be placed in closed status currently in assigned status.
9/15 KKelch scheduled per logged email from Keilah. Email logged on WO-0304116
9/15 KKelch called Keilah to schedule for 9/16 between 9 and 10 am. Left VM, sent email. Call and email logged on WO-01304116
9/12 KKelch called Keilah to schedule for 9/15 between 9 and 10 am. Left Vm. Call logged on WO-01304116
🔧Work Performed
Swapped cables on footswitch box and verified problem followed the cable normally connected to footswitch activation actuator. Replaced actuator part # 15689. All functions of the table work and unit meets manufacturers standards.
🔍Technician Findings
The calf actuator will not work - Found all functions worked except footswitch would not go up or down.
⚠️Problem Description
Model: 087
SN: 812868
Description of Equipment Issue(s) and Service(s) Requested:
Case 118814 – I will need to speak w FT before they go onsite. The calf actuator will not work. I would like to do
trouble shooting prior to parts being replaced as I’m not sure what the failure is. Please FaceTime me on an iPhone so I
can see the issue. This is a must. This could be a disconnected cable, a defective control box, a defective actuator, or a
hand control issue and I want to ensure we do what is necessary, but not more. Parts tracking # is 88423820394..
Return parts label included – tracking 884303755843. FT is expected to return parts to MPI. They can be left on the
dock at facility labeled and packaged, by FT. Please. Sign off sheet will need to be sent to MPI at the time of repair
completion. Thank you!
Part Number(s), Descriptions, and Quantities of Parts Suggested and/or Shipped (Include Ship Dates & Tracking Numbers)
Teri Masoner, 816-591-5567 cell, tmasoner@medicalpositioning.com.
***Products can be viewed at wwwmedicalpositioning.com***
Email:Re: Scheduling for repairs needed at UC Health
Completed Sep 18, 2025
From: Eric.Schafer@uchealth.org
To: Keilah.Perez@uchealth.org, kaitlyn.kelch@emsar.com
Cc: Katherine.Albrecht@uchealth.org, Britt.Smyth@uchealth.org
This is the first time you received an email from this sender (Eric.Schafer@uchealth.org). Exercise caution when clicking links, opening attachments or taking further action, before validating its authenticity. Secured by Check Point Next Tuesday works great. Do we have a list of the beds that have been repaired already?
Also, Keilah, did you r
📅
WO-01304124 - Repair - UltraScan Versa Premier Table - Aurora - UC Health Colorado Hospital
Sep 16, 2025 11:00am – 2:00pm
Work Order: WO-01304124
Account: MEDICAL POSITIONING INC.
Customer Case Number: 118814
Location: UC Health Colorado Hospital
Address: 12605 E 16th Ave, Aurora, CO 80045
Manufacturer: MPI
Model #: UltraScan Versa Premier Table
Serial #: 812868
Contact Name: Keilah Perez
Contact Phone: 720-553-5485
Co