FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
12/01/2025 Cever received email from Santos, Louis Vernon confirming Dec 8th to begin Service
12/01/2025 Cever sent email to Santos, Louis Vernon with date options for Dec 8th
11/25/2025 MHamersley - recd email from Louis Santos confirming "We have everything on the list except for TAB24180028SA. I believe it could be a typo for we have TAB24180029SA. You will need Healthtrust credentialed to enter the hospital." ... 11/25/2025 MHamersley - called 281-440-2235, spoke to Louis, POC, sending email to confirm devices; Location address is correct; email: louisvernon.santos@gehealthcare.com
🔧Work Performed
Template Name: Code 1M_screw_and_door
Template Description: Code 1M_screw_and_door
Problem Found: FMI32097_Heater door replacement.
Action Taken: FMI 32097 completed. Inspected the Heater doors. Shoulder screw disengaged and heater doors found to be damaged. Replaced both heater doors as per Service Manual Addendum (5971733) and screws tightened as per instructions.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Inspection complete. Doors replaced. Doors provided by GE.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01315408
Account: GE Healthcare FMI
Customer Case Number: 1-556489386177
Location: HCA HOUSTON HEALTHCARE NORTHWEST
Address: 710 Cypress Creek Pkwy, HOUSTON, TX 77090
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24180027SA
Contact Name: Louis Santos