FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
12/01/2025 Cever received email from Santos, Louis Vernon confirming Dec 8th to begin Service
12/01/2025 Cever sent email to Santos, Louis Vernon with date options for Dec 8th
11/25/2025 MHamersley - recd email from Louis Santos confirming "We have everything on the list except for TAB24180028SA. I believe it could be a typo for we have TAB24180029SA. You will need Healthtrust credentialed to enter the hospital." ... 11/25/2025 MHamersley - called 281-440-2235, spoke to Louis, POC, sending email to confirm devices; Location address is correct; email: louisvernon.santos@gehealthcare.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Inspection complete.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01316765
Account: GE Healthcare FMI
Customer Case Number: 1-556562136949
Location: HCA HOUSTON HEALTHCARE NORTHWEST
Address: 710 Cypress Creek Pkwy, HOUSTON, TX 77090
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24180044SA
Contact Name: Louis Santos
Co