FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
12/29/2025 EMartin rcvd email confirmation from Ron replied FSE onsite Mon 01/05/25 to complete devices
12/29/2025 MHamersley - called 561-882-2725, spoke to Ron, Bio Med Supervisor; offered Monday, January 5, 2026 for FSE to be onsite; sent email: ronald.doropan@tenethealth.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01325104 - FCO - Giraffe OmniBed Carestation - WEST PALM BEACH - St. Mary's Medical Center
Jan 5, 2026 8:00am – 5:00pm
Work Order: WO-01325104
Account: GE Healthcare FMI
Customer Case Number: 1-556721435977
Location: St. Mary's Medical Center
Address: 901 45TH ST, ,, WEST PALM BEACH, FL 33407-2413
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TABB70724
Contact Name: Ron Doropan
Contact P
JPG
TABB70724_Product Label Pic
Jan 25, 2026
JPG
TABB70724_Doors
Jan 25, 2026
JPG
TABB70724_Screw
Jan 25, 2026
Timeline
📞
Case Linked Dec 11, 1:31 PM
Case Number:00551020
Subject:GE FMI 32097 - St. Mary's
Status:Closed
📋
▶
2
Work Order Created Dec 11, 2:16 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Kimberly Hardison
From:Entered
To:Ready for Scheduling
Duration in Previous:0m
⚠️9.0 business days
✅
Status: Ready for Release by Erin MartinDec 23, 3:28 PM