FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
12/23/2025 EMartin called 229.603.9100 / Raven Toney raven.toney@trimedx.com goes to google phone number confirmed at least 20 devices onsite confirmed onsite tentaively to start devices 01/19/26 sent confirmation to madison.giggey@trimedx.com POC onsite Madison 229.288.2627 confirmed address
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01327505
Account: GE Healthcare FMI
Customer Case Number: 1-556796733216
Location: PUTNEY MEMORIAL HOSPITAL INC
Address: 417 3rd Avenue West,, Albany,, GA 31701
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24450002SA
Contact Name: Madison Giggey
Contac