FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R3.4 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed
all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries /
Batteries installed by customer and dispose of old batteries
according to local regulation”.
Verification Test:
“Performed battery charging test - verified after 10 minutes, the scan
time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”."
🔍Technician Findings
FMI 32097
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01327947 - FCO - Giraffe OmniBed Carestation - WASHINGTON - MEDSTAR WASHINGTON HOSPITAL CENTER
Jan 20, 2026 7:00pm – 7:00pm
Work Order: WO-01327947
Account: GE Healthcare FMI
Customer Case Number: 1-556819633707
Location: MEDSTAR WASHINGTON HOSPITAL CENTER
Address: 110 IRVING ST NW, WASHINGTON, DC 20010
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24240076SA
Contact Name: Shila Shrestha
C