MI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay12.3 days stuck in "Ready for Scheduling"
⏱️
Total Delay28.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
12.3d / 1.0d SLA
Ready for Release
7.4d
Scheduled
4.0d / 2.0d SLA
Closed
5.5d / 1.0d SLA
Ready to Bill
14.7d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
01/05/2026 ALuman - dispatch to Luis Iser to go onsite Tuesday 1/6. Luis has been in contact w/ POC Mark Walters about onsite visit
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
01/08/2026 LIser- FMI32097 implemented.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01324186 - FCO - Giraffe OmniBed Carestation - SOUTH MIAMI, - SOUTH MIAMI HOSPITAL
Jan 6, 2026 8:00am – 5:00pm
Work Order: WO-01324186
Account: GE Healthcare FMI
Customer Case Number: 1-556670064949
Location: SOUTH MIAMI HOSPITAL
Address: 6200 SW 73RD ST, SOUTH MIAMI,, FL 33143
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24370063SA
Contact Name: Mark Walters
Contact Phone: D
JPG
TAB24370063SA_Product Labe Pic
Jan 8, 2026
JPG
TAB24370063SA_Screw
Jan 8, 2026
JPG
TAB24370063SA_Doors
Jan 8, 2026
Timeline
📞
Case Linked Dec 5, 10:34 AM
Case Number:00549213
Subject:GE FMI 32097 - South Miami
Status:Closed
📋
▶
2
Work Order Created Dec 5, 4:35 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Kimberly Hardison
From:Entered
To:Ready for Scheduling
Duration in Previous:0m
🚨12.9 business days
✅
Status: Ready for Release by Erin MartinDec 23, 3:28 PM